Microscope
The microscope is the primary instrument of the cytogeneticist. Mastery of its optical principles is required to transform faint chromosomal structures into analyzable data
Microscope Types
-
Brightfield
- Use: Standard routine analysis (G-banding)
- Principle: Light passes through the specimen. Stained chromosomes absorb light (appear dark) against a bright background
- Requirement: Requires a Green Filter to enhance the contrast of purple Giemsa stain
-
Phase-Contrast
- Use: Quality Control (QC) of unstained slides, checking mitotic index, and harvest troubleshooting
- Principle: Converts invisible differences in refractive index (density) into visible differences in brightness. Allows visualization of “wet” transparent cells
-
Fluorescence
- Use: FISH and molecular cytogenetics
- Principle: Epifluorescence. High-energy light (Excitation) hits the sample; fluorophores emit lower-energy light (Emission). The background is black. Requires specific filter cubes (DAPI, FITC, Texas Red)
Components & Functions
- Support: The heavy Stand and Stage provide stability
-
Illumination
- Lamp: Source of white light
- Condenser: Concentrates light into a cone to illuminate the specimen
- Diaphragms: Field Diaphragm controls beam width (glare); Aperture Diaphragm (Iris) controls contrast vs. resolution
-
Optics
- Objectives: The primary lenses (10x Scan, 100x Oil)
- Eyepieces: Further magnify the image (10x). Total Mag = \(100 \times 10 = 1000\text{x}\)
Achieve Optimal Resolution
Resolution (\(d\)) is the ability to distinguish two close points. \(d = \lambda / (2 \cdot NA)\). To minimize \(d\) (improve resolution):
- Decrease Wavelength (\(\lambda\)): Use a Green Filter (shorter wavelength than red)
- Increase Numerical Aperture (\(NA\)): Use Immersion Oil (\(n=1.515\)) with the 100x objective. Oil bridges the air gap, preventing light refraction/loss
- Align Light: Perform Kohler Illumination to center and focus the condenser
- Optimize Iris: Set the Condenser Iris to ~75% open to balance resolution and contrast
Maintenance & Troubleshooting
- Daily: Clean oil off the 100x lens after use (oil dissolves cement). Cover to prevent dust
-
Troubleshooting
- Blurry Image: Usually a dirty objective or slide. Clean thoroughly
- Black Field: Check light path (camera pull-rod engaged? Field diaphragm closed?)
- Drifting Focus: Tighten the tension collar on the focus knob
- Uneven Light: Condenser is off-center (Fix: Kohler)