Troubleshoot Processing
FISH is a complex multi-step assay where chemical, physical, and optical variables interact. When an assay fails, the “symptoms” observed under the microscope (e.g., no signal, high background, cross-hybridization) can be traced back to a specific failure point in the protocol. The laboratory scientist uses a troubleshooting algorithm to identify the root cause and determine if the slide can be salvaged (re-processed) or if the assay must be repeated from the beginning
Problem 1: No Signal (The “Black Slide”)
- Symptom: DAPI nuclei are visible, but no Red/Green probe signals are seen
-
Root Cause Analysis
- The Probe: Was the probe actually added? (Human error). Was the probe expired or degraded (left in light)?
-
Denaturation (The Prime Suspect)
- Under-Denaturation: Check the DAPI. If nuclei are shiny/refractile, the DNA never melted. The probe bounced off
- Fix: You can re-denature this slide. Wash off the DAPI, re-apply probe, and denature at a higher temp (e.g., \(+2^\circ\text{C}\))
-
Wash Stringency
- Over-Washing: Was the post-wash too hot (\(>74^\circ\text{C}\))? This strips the specific signal
- Fix: Cannot save this slide. Must repeat with a new slide and lower wash temp
- The Microscope: Is the shutter open? Is the fluorescence lamp (bulb) dead or old (>200 hours)?
Problem 2: Weak Signal (Faint/Grainy)
- Symptom: Signals are visible but require straining the eyes. They fade instantly (photobleaching)
-
Root Cause Analysis
- Specimen Quality: “Old” slides (stored for years) lose DNA integrity
- Drift: The optical filters or lamp alignment may be off
- Re-Processing: If a slide was processed twice, the DNA degrades, leading to weaker signals each time
-
Fix
- Increase the hybridization time (e.g., 48 hours instead of 16)
- Use a fresh probe aliquot
- Check the Antifade (expired antifade allows rapid bleaching)
Problem 3: High Background (The “Green Haze”)
- Symptom: The entire nucleus or the glass slide glows green/red. It is hard to distinguish true dots from the noise
-
Root Cause Analysis
-
Wash Stringency (Under-Washing): The post-wash was too cold or too salty. The non-specific probe wasn’t removed
- Fix: You can re-wash this slide. Remove coverslip, wash again at the correct temp, and re-counterstain
-
Drying: If the slide dried out during hybridization (chamber wasn’t humid), the probe binds irreversibly to the glass
- Fix: Unsalvageable. Repeat
-
Cytoplasm: Heavy cytoplasm traps probe
- Fix: Treat the next slide with Pepsin to digest cytoplasm before hybridizing
-
Wash Stringency (Under-Washing): The post-wash was too cold or too salty. The non-specific probe wasn’t removed
Problem 4: Cross-Hybridization (Non-Specific Binding)
- Symptom: In addition to the 2 bright target signals, there are many fainter signals on other chromosomes
-
Root Cause Analysis
- Cot-1 DNA Failure: The repetitive sequences (Alu repeats) in the probe were not blocked
- Probe Concentration: Too much probe was used
- Homology: The target gene has a “pseudogene” elsewhere in the genome with similar sequence
- Fix: Increase the stringency of the post-wash (higher temp) to strip off the weaker cross-hybridized binding while keeping the strong specific binding
Problem 5: Autofluorescence (Glowing Debris)
- Symptom: Debris or cells glow yellow/orange (visible in both Red and Green channels)
-
Root Cause Analysis
- Biological: Red blood cells, necrosis, or collagen naturally fluoresce
-
Fix
- During Analysis: Switch between filters. True signal is only visible in one channel (e.g., Red). Autofluorescence is visible in all channels
- During Prep: Use a fixative wash (Carnoy’s) to lyse RBCs more thoroughly. For tissue, use specialized “Quenching” reagents (e.g., Sudan Black) to mask autofluorescence